Water activity / Moisture

Water activity meter - Aqualab Series 4TE Duo

Water activity and moisture content measurement in 5 minutes or less.

The AquaLab Series 4TE water activity meter has internal temperature control which allows you to choose a measuring temperature between 15 and 50°C. Since water activity is temperature dependent, the AquaLab Series 4TE will allow you to expertly manage the internal temperature to precisely measure the water activity of your product.

The AquaLab series 4 can be used in:
  • Accelerated shelf-life studies
  • Isotherm studies
  • Studies determining the effects of temperature on a sample aw
  • Evaluate the effects on a sample when there are large fluctuations in ambient temperature
  • Meeting government or company regulations on temperature specific aw measurements

•0.0001 aw resolution

•Ability to restrict user access and to control which instrument settings can be changed by users

•Deal sensor block with integrated dew-point and volatile sensors in one*

•Displays water activity and moisture content simultaneously**

•Data storage (stores up to 16000 readings for later download)

•New 'flip-top' block gives easy cleaning access

•Improved sample chamber seal

•Advanced software reporting package

Sensor Types: a. Chilled-mirror dewpoint. b. Infrared temperature.

Accuracy: ±0.003 aw

Range: 0.10 to 0.95 aw

Moisture Content Precision: 0.02%

Agreement to Moisture Content Reference Method: ±0.1% to ±0.5%

Sample Dish Capacity: 7 ml recommended (15 ml full)

Resolution:

Moisture: ±0.01% mc

Water Activity: ±0.0001 aw

Balance: not applicable, none.

Measurement Speed: Less than 5 minutes.

Display: 128x64 pixel LCD display with backlighting

Results Displayed Percent moisture and water activity.

Temperature Control 15 to 50°C (±0.2°C)

Temperature Stability User-selectable range, internal thermoelectric controlled.

Test Result Memory 10,000 readings (each reading includes water activity, moisture content, temperature, time, date, operator, and sensor used)

Program Identification Alphanumeric; Programmable to display product name, lot, or product ID number.

Operating Environment 4 to 50°C (39.2 to 122°F) 0 to 90% Relative Humidity (non-condensing.)

Universal Power 110 V to 220 V AC, 50/60 Hz Less than 0.4 amps.

Data Interface RS232A compatible, 8-data bit ASCII code, 9600 baud, no parity, 1 stop bit, cable included.

Case Dimensions 26.7 x 17.8 x 12.7 cm (11 x 7.1 x 5.1 in)

Case Material Machined aluminum frame; injection molded plastic cover.

Weight 3.18 kg (6.9 lb)

Warranty Three years, factory parts & labor.

Certifications CE; AOAC Approved Method for Measurement of Water Activity

 

Water Activity and Moisture Content in Snack Foods

Water Activity and Moisture Content in Candy and Confectionery

Water Activity and Moisture Content in Cereal and Grain Products

Water Activity and Moisture Content in Coffee and Tea

Water Activity and Moisture Content in Cosmetics

Water Activity and Moisture Content in Plastics and Other Materials

Water Activity and Moisture Content in Nuts and Dried Fruit

Water Activity and Moisture Content in Pet Food

Water Activity and Moisture Content in Pharmaceuticals

Water Activity and Moisture Content in Powders and Spices

Water Activity and Moisture Content in Processed and Cured Meats

Water Activity and Moisture Content in Sauces and Dressings

Water Activity and Moisture Content in Seeds

Water Activity and Moisture Content in Tobacco

Water Activity and Moisture Content in Buildings

Request A Quote Download Brochure